Duplicate breath samples using the Dräger 7110.

Auteur(s)
Laslett, R.
Jaar
Samenvatting

The aim of this study was to determine the comparison of duplicate breath samples taken on a Dräger Alcotest 7110, Mk II. A total of 216 persons were tested with the second sample being obtained 2 to 10 minutes after the first sample had been analysed. The mean difference between duplicate samples is 0.003 grams of alcohol in 100 milliliters of blood. The minimum difference was 0.000 grams of alcohol in 100 milliliters of blood and a maximum difference 0.015 grams. The failure rate between 0.020 and 0.049 (n=79) is zero with a standard deviation of 0.0014. The failure rate where the BAC is >0.050 (n=101) is <1 % with a standard deviation of 0.0033. 36 subjects were less than 0.020 grams of alcohol in 100 milliliters of blood. Duplicate breath samples can be successfully obtained on a Dräger Alcotest 7110 Mk II instrument coupled with a good quality assurance programme.